Person Responsible for Regulatory Compliance under Article 15 of the IVDR

In the realm of In-vitro diagnostic Medical devices, regulatory compliance is non-negotiable. The European Union In Vitro Diagnostic Regulation (IVDR) have set stringent standards to ensure the safety and efficiency of medical devices. A pivotal role introduced by these regulations is the Person Responsible for Regulatory Compliance (PRRC). The PRRC is tasked with ensuring that all regulatory requirements are met, from product development through to post-market surveillance. This role is essential for maintaining the quality and safety of medical devices, and compliance with these regulations is mandatory for companies operating within the EU.

However, for many companies, especially small to medium-sized enterprises (SMEs), the challenges associated with maintaining an in-house PRRC can be daunting. This is where outsourcing the PRRC role comes into play. Article 15 of the IVDR allows the utilisation of external PRRCs for micro and small enterprises. IVDeology have a proven track record in providing expert regulatory expertise to SMEs.

Company categoryStaff headcountTurnoverorBalance sheet total
Medium-sized< 250≤ € 50 m≤ € 43 m
Small< 50≤ € 10 m≤ € 10 m
Micro< 10≤ € 2 m≤ € 2 m

Source: SME definition – European Commission (europa.eu)



Access to Specialised Knowledge

One of the most significant benefits of outsourcing the PRRC role is the access to specialized expertise. Regulatory compliance is a complex and ever-evolving field, requiring a deep understanding of both the regulations and the specific products being developed. Outsourcing to a third-party provider ensures that the PRRC has the necessary experience and knowledge to navigate these complexities. These experts are typically well-versed in the intricacies of the IVDR, and they stay abreast of the latest developments in the regulatory landscape. This level of expertise can be particularly advantageous for companies developing innovative or highly specialized products, as it ensures that all regulatory requirements are met to the highest standards.



Saving Time and Money in the Long Run

Hiring and training an in-house PRRC can be a costly endeavour. The recruitment process itself can be expensive, and once a suitable candidate is found, they need to be trained and continuously educated on the latest regulatory requirements. Additionally, the salary for an experienced PRRC can be substantial. By outsourcing this role, companies can reduce these costs significantly. Third-party providers typically offer their services on a contract basis, which can be more cost-effective than employing a full-time, in-house PRRC. This approach allows companies to allocate their financial resources more efficiently, investing in other critical areas of their business such as research and development, marketing, or expanding their product portfolio.

Our aim is to work with our clients to develop a succession plan for the eventual internal replacement of our virtual PRRC. We will train and mentor the identified internal team members so build the experience required to meet the requirements of Article 15.

Letting Experts Handle Compliance

Regulatory compliance, while essential, is often not a company’s core activity. For many organizations, particularly SMEs, the primary focus is on developing innovative products, enhancing customer experiences, and growing their business. We understand that some organisations may not want to build their regulatory team internally, so we always try to build long term relationships with our customers for continuous support. By outsourcing the PRRC role, companies can allow their internal teams to focus on these core activities. This not only enhances productivity but also ensures that the regulatory compliance function is handled by experts. This separation of duties can lead to better outcomes in both regulatory compliance and overall business performance.

“Regulatory compliance is the responsibility of everyone in the organisation, the PRRCs core responsibility is to observe, reflect and advise on the overall compliance state of the organisation to senior management”

Adapting to Changing Needs

The regulatory landscape is constantly changing, and the demands placed on the PRRC can vary significantly over time. Outsourcing the PRRC role offers companies the flexibility to scale their regulatory compliance efforts up or down as needed. For instance, during periods of intensive product development or regulatory scrutiny, a company may require more comprehensive compliance support. Conversely, during quieter periods, the level of support needed may be reduced. Outsourcing allows companies to adapt to these changing needs without the long-term commitments associated with employing a full-time, in-house PRRC. This flexibility can be particularly beneficial for companies with fluctuating workloads or those operating in highly dynamic industries.

Ensuring Compliance and Avoiding Penalties

Non-compliance with regulatory requirements can result in severe penalties, including fines, product recalls, and even the suspension of business operations. The PRRC plays a crucial role in mitigating these risks by ensuring that all regulatory requirements are met and that any potential issues are addressed promptly. Outsourcing the PRRC role to a reputable third-party provider can enhance this risk mitigation process. These providers have the experience and resources to conduct thorough compliance assessments, implement robust quality management systems, and respond swiftly to any regulatory changes or issues. This proactive approach can help companies avoid the costly consequences of non-compliance and maintain their reputation in the market.

Staying Ahead of Regulatory Changes

The IVD industry is characterized by rapid technological advancements and frequent regulatory updates. Staying ahead of these changes is critical for maintaining compliance and achieving business success. Outsourcing the PRRC role to a third-party provider ensures that companies have access to the latest regulatory knowledge and best practices. These providers invest in continuous education and professional development to stay current with regulatory changes. By leveraging their expertise, companies can continuously improve their compliance processes, adapt to new regulations seamlessly, and maintain a competitive edge in the market.

Giving you a Strategic Advantage

Outsourcing the PRRC role offers numerous benefits for companies operating in the medical device industry. From accessing specialized expertise and achieving cost savings to enhancing focus on core activities and adapting to changing needs, outsourcing provides a strategic advantage. By partnering with a reputable third-party provider, companies can ensure that their regulatory compliance efforts are handled by experts, mitigating risks and driving continuous improvement.

In an industry where compliance is critical to success, outsourcing the PRRC role can be a game-changer, allowing companies to navigate the complexities of regulatory requirements with confidence and achieve their business goals.

For further information on how we can support you, please contact us.

Written by Stuart Angell, Co-founder and Director of IVDeology, part of TEAM PRRC

Posted on

Near-patient Testing – Is it the Future for Diagnostics?

There has been a lot of discussion surrounding the focus of treatment within primary care.  The NHS published guidance in August 2023 for integrating point of care IVD’s (NHS England » Integrating in vitro point of care diagnostics: guidance for urgent community response and virtual ward services).  The Labour Party’s recent manifesto for the 2024 General Election stated that “Labour’s reforms will shift our NHS away from a model geared towards late diagnosis and treatment, to a model where more services are delivered in local communities” and “The National Health Service needs to move to a Neighbourhood Health Service, with more care delivered in local communities to spot problems earlier. To achieve this, we must over time shift resources to primary care and community services.”  (Change Labour Party Manifesto 2024).  Whilst this is good news for patients, it is not as simple as changing the location of the testing – there will be implications for the IVD industry in order to meet these proposals.

The In Vitro Diagnostic Medical Devices Regulation (Regulation (EU) 2017/746, ‘IVDR’) defines a “device for near-patient testing” as “any device that is not intended for self-testing but is intended to perform testing outside a laboratory environment, generally near to, or at the side of, the patient by a health professional.”

During Design & Development, IVD manufacturers have to determine the use environment and end users within their intended purpose.  It is then their responsibility to demonstrate that the device is safe and effective when used in these environments by these intended end users and therefore there will be a number of additional requirements that need to be considered for devices intended for near-patient testing.

Performance Requirements:

The performance testing conducted will need to demonstrate that effective and reliable testing outside of the traditional controlled laboratory environments.  Considerations such as the following will need to be considered:

  • Use Environment Testing:

Annex I section 9.4 states “The characteristics and performances of the device shall be specifically checked in the event that they may be affected when the device is used for the intended use under normal conditions:

(b) for devices for near-patient testing, performances obtained in relevant environments (for example, patient home, emergency units, ambulances).

Testing of the device will therefore need to be conducted in the intended use environments. Manufacturers will need to show that the test can be used reliably in the environments that they indicate that the test can be used in. This could be, for example, doctor’s surgeries, A&E departments, patient’s homes or in ambulances.  Any conditions specific for these environments would need to be considered e.g. vibrations or temperature fluctuations for devices intended to be used on an ambulance.

  • Usability Studies:

Annex I section 19.1 states that “Devices intended for self-testing or near-patient testing shall be designed and manufactured in such a way that they perform appropriately for their intended purpose taking into account the skills and the means available to the intended user and the influence resulting from variation that can be reasonably anticipated in the intended user’s technique and environment. The information and instructions provided by the manufacturer shall be easy for the intended user to understand and apply in order to correctly interpret the result provided by the device and to avoid misleading information. In the case of near-patient testing, the information and the instructions provided by the manufacturer shall make clear the level of training, qualifications and/or experience required by the user.

It is important for manufacturers to have designed their device in such a way as to ensure that the intended end users can successfully use the device.  Therefore it is vital that manufacturers conduct usability studies with their target end users who are often not laboratory trained personnel.  This can then be used to demonstrate that consistent results can be obtained by these target end users, determine if the instructions provided with the device are adequate and help to identify if any training is required for the end users before the device can be used reliably.

Labelling Requirements:

The IVDR has also introduced specific labelling requirements for NPT (near patient testing) devices, including:

  • Devices labelled as Near-Patient Testing: The device label must indicate that the device is for near-patient testing.  Although there are currently no symbols for this within ISO 15223-1:2021 Medical devices – Symbols to be used with information to be supplied by the manufacturer, MedTech Europe has provided some suggested symbols that can be used to indicate near-patient testing (New IVD symbols for compliance with the IVDR – MedTech Europe).
  • Individual Instructions for Use: Each individual device must be accompanied by its own instructions for use (IFU).  For devices for professional use within a laboratory setting, if multiple devices were supplied then a single copy of the IFU could be provided if agreed by the purchaser.  This is not allowed for devices intended for near-patient testing.
  • Paper-Based Instructions: According to Annex I 20.1(f) of the IVDR, the instructions for use must be provided in a physical paper format for near-patient tests, and, unlike laboratory based professional use devices, cannot be provided in electronic format.
  • Language Requirements: The languages that the Member States require for the device label and instructions for use for near-patient testing may be different to those for professional use only tests.  This may add more translation costs on to manufacturers to access different markets.

The labelling provided must be appropriate to the device, its intended use, and the technical knowledge, experience, education, or training of the intended users.  This will need to be considered by manufacturers when designing the labelling and tested during usability studies.

Notified Body Assessment

Whilst devices for near-patient testing are classified in their own right according to Annex VIII rule 4(b), the notified body assessment is slightly different.  For Class B & Class C devices, the technical documentation of all devices for near-patient testing has to be assessed rather than the notified bodies sampling one technical file per generic device group or device category. Where a manufacturer has a number of devices for near-patient testing the increase of upfront cost to have their devices assessed will need to be considered.

Final thoughts

Whilst near-patient testing seems like a real win for patients and the direction of travel that the diagnostic industry is heading, this does provide some challenges for manufacturers.  The additional burden to demonstrate that the device is effective and reliable when used outside of a laboratory setting and the potential increased upfront costs of conformity assessment is something that needs to be considered before being able to place the device on the market.  For Great Britain, although the UK MDR 2002 does not specifically call out devices for near-patient testing currently, the indications from the MHRA on the future regulations is that it will be similar to the IVD Regulation with the new Essential Requirements being based on the General Safety & Performance Requirements.  This is likely therefore to mean that the additional requirements for these types of devices will also be required here.  However, if done correctly, near-patient testing will enable quicker diagnoses for patients and hopefully therefore better patient outcomes, which is ultimately what the IVD industry wants to support.

If you’d like to discuss near-patient testing or any of the compliance services that come along with it, from design and development to regulatory services, you can speak to us by dropping an email to [email protected] or book time with us via this link for when best suits you

Posted on

Performance Evaluation: The critical component in bringing your product to market and keeping it there

Abingdon Health and IVDeology Ltd have decades of  In-Vitro Diagnostic (IVD) regulatory compliance experience where our teams support your entire product journey from ‘cradle-to-grave’ to ensure you are getting your product to market in the quickest timeframe possible, reducing cost and strain, and keeping it compliant through its lifetime.

Abingdon Health also offer full-service contract development and manufacturing for lateral flow assays, bringing your idea to commercial success, with the benefit of an integrated regulatory and quality approach.

In our latest blog, Candice Vendettuoli Head of RAQA at Abingdon Health covers the importance of getting your Performance Evaluation (PE) right to streamline your route to market, and ensuring compliance to keep it there.

What is Performance Evaluation?

The In Vitro Diagnostic Regulation (IVDR) 2017/746, which entered into force in May 2017 and applied in May 2022, has introduced significant changes to the way in vitro diagnostic (IVD) medical devices are regulated in the European Union. One of the key aspects of this regulation are the new requirements for documenting the performance evaluation of IVD medical devices using a prescriptive document structure mandated within the Regulation.

Performance evaluation under IVDR is expected to be a continuous process throughout its entire lifecycle of the device. This process is crucial for ensuring that the device meets upon entry to the market, and continues to meet, the intended clinical benefits and safety as claimed by the manufacturer.

The mandated documents should be written to provide a comprehensive and structured narrative for the reviewer giving a clear and logical explanation of how the device was developed, verified and validated against the intended use/purpose claimed by the manufacturer. These documents are a requirement of the Technial Documentation described in Annex II and forms an essential part of the submission to the Notified Body

The mandated documents, unless they can justify why such studies are not applicable are as follows:

Performance Evaluation Plan

Ideally written during the early development of the device and updated regularly, this document has content prescribed within Annex XIII section 1.1 of the Regulation. Manufacturers are required to establish and regularly update the performance evaluation plan that outlines the device’s characteristics and performance, as well as the process and criteria used to generate the necessary clinical evidence.

Scientific Validity

The concept of scientific validity under the In Vitro Diagnostic Regulation (IVDR) 2017/746 is a cornerstone in the performance evaluation of in vitro diagnostic (IVD) medical devices. It refers to the association of an analyte with a clinical condition or physiological state, which must be substantiated with a medical-scientific rationale evidenced through a systematic literature search

Analytical Performance

Analytical performance refers to a device’s ability to accurately and reproducibly measure an analyte, marker, or molecule, which is a strictly technical performance without the need for correlation with a targeted pathology.  There are analytical performance characteristics mandated within Annex I section 9.1(a) including assessing the accuracy, sensitivity and specificity,

Clinical Performance

Clinical performance is defined as the ability of a device to yield results that are correlated with a particular clinical condition, physiological or pathological process, or target population and intended user. Manufacturers must demonstrate clinical performance through one or more of the following:

  • Clinical performance studies, carried out according to the IVDR requirements on clinical performance studies described in Articles 57-77, Annex XIII section 2 and, if applicable, Annex XIV for studies other than those using leftover samples
  • Scientific peer-reviewed literature on the device under evaluation, or
  • Published routine diagnostic testing.

Performance Evaluation Report

The report (also known as a ‘PER’) provides a summary of the clinical evidence collected through the previous reports. An assessment can then be made against the current state of the art in diagnostics and medicine that a positive benefit-risk ratio of using the device for its intended purpose has been met and then all data has been collected.

This rigorous approach ensures the reliability and effectiveness of in vitro diagnostic devices within the European Union, with the primary aim of protecting public health by requiring high levels of safety and performance of these devices to be evidenced.

For manufacturers, understanding and adhering to the IVDR’s performance evaluation requirements is vital for successful market introduction of their IVDs in the European Market. It involves a comprehensive understanding of the general safety and performance requirements (GSPR), as well as the specific guidelines on performance evaluation stipulated in Article 56 of the IVDR.